{"ob_id":43195,"uuid":"a6f65a61de0f44d7b8c3ba24cfb72cb7","title":"National Physical Laboratory ANSTO radon A","abstract":"The Australian Nuclear Science and Technology Organisations (ANSTO) 1500L low-level radon monitor provides a unique technology for highly sensitive measurements\r\n     of atmospheric radon (< 100 mBq m-3) and is recognised by the World Meteorological Organisation (WMO) as the best in the world for global and atmospheric\r\n     compositional baseline studies. The ANSTO dual-flow-loop two-filter works by removing all ambient radon (222Rn) and thoron (220Rn) progeny in the sampled air\r\n     from the airstream using a filter before the air passes into a large delay volume to allow for the decay of the short-lived isotope 220Rn. The sample air in the\r\n     detector is passed through a flow loop at a flow rate designed to exchange the delay volume's air in approximately 20 minutes. This allows time for new radon\r\n     progeny to form. A second flow loop circulates the volume of the delay chamber through the second filter (a low-impedance 625 mesh stainless steel screen) to\r\n     collect all 218Po progeny. The newly formed progeny are collected on the second filter and their decays are counted with a ZnS-photomultiplier system. Using the\r\n     flow rate and the count rate the atmospheric 222Rn concentrations can be calculated. This analyser has been used at Heathfield atmospheric monitoring station.","keywords":"","identifier_set":["https://catalogue.ceda.ac.uk/api/v2/identifiers/13191/?format=json"],"instrumentType":"","subInstrument":[],"responsiblepartyinfo_set":["https://catalogue.ceda.ac.uk/api/v2/rpis/205455/?format=json"]}