Dataset
Continuous measurements of atmospheric carbon dioxide (CO2) and oxygen (O2) at Heathfield Tower 2021-2022
Abstract
Continuous measurements of atmospheric carbon dioxide (CO2) and oxygen (O2) made at at Heathfield Tower, East Sussex, UK for the DARE-UK (Detection and Attribution of Regional greenhouse gas Emissions in the UK) project.
The measurement system was installed in June 2021 and measurements are ongoing. There were two inlet lines 100 meters above ground level, going up the tower, automatic switching between inlet lines every 2 hours. Data were sampled every 2 minutes, and gaps are present due to routine running of calibration cylinders, maintenance & removal of data with known technical issues.
Atmospheric CO2 was observered using a Non-Dispersive InfraRed (NDIR) CO2 analyser from Siemens Corp., model Ultramat 6E (Units: ppm), whilst atmospheric O2 was measured using a dual lead fuel cell O2 analyser, Oxzilla from Sable Systems International Inc. (units: per meg). These CO2 and O2 measurements were then combined to calculate Atmospheric Potential Oxygen (APO) in units of per meg .
The data contain quality flags as follows:
Flag 1 : Good data
Flag 2 : From 19Jun2021 to 31Aug2021 there was no air conditioner in the laboratory, so the oxygen data has been adjusted using the room temperature.
Calibration scales:
CO2 data were transferred onto the WMO NOAA X2019 scale using Equation 6 in Hall et al., 2021, doi.org/10.5194/amt-14-3015-2021
O2 data are on the Scripps SIO S2 scale.
Details
Previous Info: |
No news update for this record
|
---|---|
Previously used record identifiers: |
No related previous identifiers.
|
Access rules: |
Public data: access to these data is available to both registered and non-registered users.
Use of these data is covered by the following licence(s): http://www.nationalarchives.gov.uk/doc/open-government-licence/version/3/ When using these data you must cite them correctly using the citation given on the CEDA Data Catalogue record. |
Data lineage: |
Data were collected and processed by the project team before uploading to the CEDA archive |
File Format: |
Data are BADC-CSV formatted.
|
Process overview
Instrument/Platform pairings
Weybourne Atmospheric Observatory: Siemens Ultramet 6E & Sable Systems Oxzilla II | Deployed on: Heathfield Tower, East Sussex |
Output Description | None |
- units:
- var_id: 6
- long_name: APO.permeg
- units:
- var_id: 7
- long_name: APO.sd.permeg
- units:
- var_id: 2
- long_name: CO2.ppm
- units:
- var_id: 3
- long_name: CO2.sd.ppm
- units:
- var_id: 4
- long_name: O2.permeg
- units:
- var_id: 5
- long_name: O2.sd.permeg
- units:
- var_id: 1
- long_name: date.time
- units:
- var_id: 8
- long_name: flag
Co-ordinate Variables
Temporal Range
2021-06-01T00:00:00
2022-07-31T00:00:00
Geographic Extent
50.9767° |
||
0.2306° |
0.2306° |
|
50.9767° |